Herunterladen Diese Seite drucken

limoss MD200 Bedienungshinweise Seite 11

Vorschau ausblenden Andere Handbücher für MD200:

Werbung

Neu anlernen /link up the handset
To re-learn a radio transmitter, please proceed as follows.
Error Case 1 in detail
: Both transmitter control both beds at the same
time.
1. Lay the hand transmitter 1 to bed 1
2. Lay the hand transmitter 2 to bed 2
3. Separate bed 1 from the mains voltage and disconnect the optional
emergency lowering.
4. Separate bed 2 from the mains voltage and disconnect the optional
emergency lowering.
5. Turn on the voltage of bed 1 again after the switching power supply
LED is off. (After round about 20 seconds)
6a. Immediately both S1 and S2 keys from the handset transmitter 1,
simultaneously press and hold
6b. so long until the controller issues a long beep.
7. The head motor runs at the same time up, down, up. This is the
feedback that your system is connected.
8. A second time at the transmitter 1 is still being taught, it starts again
with step 3
9. All steps starting from step 4 must now be performed exactly for bed 2
again a secound time.
(The transmitter must be taught to disconnect and release from the
system. Both memory slots are overwritten and resetet now.)
Error Case 1:
HS1 adjust both beds = HS1 tech 2x to bed 1
HS2 adjust both beds = HS2 tech 2x to bed 1
Error Case 2:
HS1 adjust both beds
HS2 adjust bed 2 = HS2 teach2x to bed 2
Error Case 3:
HS1 adjust bed 1
HS2 adjust both beds = HS1 tech 2x to bed 1
-11-

Werbung

loading

Verwandte Produkte für limoss MD200

Diese Anleitung auch für:

Md201