2.6. Data Interface Option D

By means of the cable supplied the concentrator
can be conntected to a computer with SB-interfa-
ce. This makes the concentrator's service easier.
Apart form that the interface serves as a connec-
tion to a low consumption oxygen monitor.

2.7. Ionization Option I

To use the ionization option I unscrew the hose
nozzle and connect the ionization supplement in
its place. Connect the plug to the relevant bush.
Put the oxygen mask on the ionziation head.
We recommend to apply an anti-static spray to
the mask every time before it is used. Additional-
ly an anti-bacteria filter can be inserted.
Do not use a humidifyer during ionization.
The ionization can be activated by pressing I/O
on the remote control supplied. The treatment in-
tervals are set with the buttons marked
in 5 minutes steps between 5 and 95 minutes.
The treatment begins by pressing Start. The re-
maining treatment time is indicated in the dis-
play. The treatment can be interrupted by pres-
sing Pause and continued by pressing Start.
An acoutsic signal indicates the end of the tre-
atment and End of Treatment appears in the dis-
play.

4. Troubleshooting Guide

Symptom
Device does not run
Display: MAINS
Interval Sound
Device is running
Display: 02
82%
Compressor off
Display: O2
72%
Device is running
Display: 02 Flow too low
The detachable ionization head should be clea-
ned once a week or when used by a different pati-
ent (see also 5. Cleaning).

3. Diagnosis

An extensive autodiagnosis system has been im-
plemented into the oxygen concentrator (PRECI-
SE 6000).
By pressing the button next to the oxygen out-
let on the right the diagnosis is called up.
The displays can be called up by pressing the
button.
Display 1: Service Telephone
Display 2: Software Version,
and
Display 3: Date, Time
Display 4: Oxygenconcentration at Option M
Display 5: Pressure Evaluation
Display 6: Working Hours Counter
Display 7: Interval Time
Display 8: 1. Error Memory Bank
Display 9: 2. Error Memory Bank
Display 10: 3. Error Memory Bank
Cause
No line voltage
Defective fuse
Flow too high
Oxygen concentration
less 72%
Flow regulator turned off
Oxygen hose break
15
Equipment Version
Remedy
Check if mains cable is
connected correctly
In case check main fuse
Call service
Set to correct flow
Turn device on and off
call service
Set to correct flow
Check hose