Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Specifications - ConMed C7115 Bedienungsanleitung

Apex-handsteuerung
Inhaltsverzeichnis

Werbung

11. Steam sterilization is safe and effective and has no contraindications for sterilizing this equipment. Sterilize per the
following table.
Method
Steam (wrapped)
Steam (wrapped)
TROUBLESHOOTING
Symptom
Remote does not operate

TECHNICAL SPECIFICATIONS

Classification:
Protection Ingress of Fluids:
Mode of Operation:
Remote Cable Length:
Weight:
ENVIRONMENTAL REQUIREMENTS
Environmental Conditions
Temperature:
Relative Humidity:
Atmospheric Pressure:
Cycle
Minimum Temperature
Pre-vacuum
Gravity
Possible Cause
Cord(s) not connected securely.
Door is not closed.
Remote is faulty.
Remote button stuck down.
Class I (intended for connection to Class I surgical console). Type rating of
applied part is dependent on surgical console.
IPX0 (ordinary)
Continuous
11 ft. (3.35 m)
12.4 oz. (352 g)
Operating
30%
Non-Condensing
Minimum Exposure
270°F (132°C)
250°F (121°C)
Corrective Action
Storage and Transport
75%
10%
3
Minimum Dry Time
4 minutes
30 minutes
Securely connect Remote to console.
Close door on remote body.
Return for service.
Open remote door and reseat button
on door.
93%
Non-Condensing
8 minutes
8 minutes

Werbung

Inhaltsverzeichnis
loading