Outlet Pressure
Inlet SO
2
Outlet SO
2
Outlet Temperature
Note: The upper alarm limit must always be greater than the lower alarm limit for any
given parameter.
11. AUDIBLE AND VISUAL SIGNALS
Indicator Light
Not illuminated
Green/Yellow Flash 1x
Green
Yellow, solid
Yellow, flashing
12. MAIN SCREEN TILES
Display
Blue inlet pressure
Red outlet pressure
Gray delta pressure
Blue inlet oxygen saturation
Red outlet oxygen saturation
Red outlet temperature
Blue, white, or red color numerical
value
Black numerical value with yellow
solid or yellow flashing background.
Arrow up.
50 / 300 mmHg
50.0 / 90.0 %
95.0 / n/a %
35.0 / 37.5 °C
Volume
Signals
-
-
-
1x 2 beeps, 1x 3 beeps
-
-
60 - 68 db(A)
2 beeps repeating
every 30 seconds
66-71 db(A)
3 beeps repeating
every 7.5 seconds
29
0 mmHg
750 mmHg
40.0 %
100 %
40.0 %
100 %
20.0 °C
40.0 °C
Status
Device off or measured
parameter alarms inactive
Power on test sequence
Device on, measured
parameter alarms active,
no alarm condition
Low priority alarm
Medium priority alarm
Details
Pressure at the oxygenator inlet barb
Pressure at the oxygenator outlet barb
Difference between the outlet and inlet
membrane pressures
Blood oxyhemoglobin saturation at the
oxygenator inlet barb
Blood oxyhemoglobin saturation at the
oxygenator outlet barb
Blood temperature at the oxygenator
outlet barb
Reading within alarm limits
Reading exceeding upper alarm limit