Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting; Possible Cause; Corrective Action - AIDIAN QuikRead go Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5 TROUBLESHOOTING

Error message / Troubleshooting
QuikRead Instrument go does not start.
Touch panel on screen is not functioning
properly.
Instrument alarm sounds cannot be heard.
QuikRead go Instrument displays error messages
and guides the user in case it detects errors. Fol-
low the instruction displayed and see the trouble-
shooting table in this instruction for use and in
QuikRead go kit instruction for use.

Possible cause

Power unit is not connected.
The instrument has an electronic malfunction.
Touch calibration is not correct
i.e. the active area is not under the button.
Touch panel is not responding at all.
Volume is set at a low level.
Instrument sound system has a malfunction.
Contact your local supplier in case of needed sup-
port or a repair requirement.

Corrective action

Connect power unit and retry.
Contact customer service.
Calibrate the touch screen according to the pro-
cedure described in the Maintenance chapter.
Contact customer service.
Set volume according to the procedure de-
scribed in the Section "Personal settings".
Restart QuikRead go Instrument. If the problem
persists, please contact customer service.
TROUBLESHOOTING
29

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

133893149915

Inhaltsverzeichnis