Damage to the power cord may cause fire or shock hazard. When unplugging the power cord,
hold it by the plug only and remove the plug carefully.
If the power cord needs to be replaced, replace it only with another Codonics power cord
manufactured specifically for your imager's power configuration.
If the imager is smoking or making unusual sounds, power off and unplug the imager
immediately.
Do not insert foreign objects of any kind into the imager; doing so can constitute a safety hazard
and cause extensive damage.
Do not place any liquid containers on the imager. If, for some reason, liquid seeps into the
imager, power off the imager and unplug the power cord from the source outlet. If used
without corrective measures, the imager may be damaged.
Do not use the imager near flammable gases.
With the imager top cover open or the receive trays removed, touch only those internal
components that are colored green (shown in the following figure). Remove rings, ties, jewelry,
and other items, and tie back hair, so that they do not fall into or get caught in the imager.
Internal Components That Are Colored Green (called out in the illustration) Are Safe to Touch

Location Precautions

The imager's operating ambient temperature range is 15–30ºC
(59–86ºF), with a relative humidity of 10%–80%.
If the imager is moved quickly from an extremely cold place to a warmer one, condensation is
likely to form. Do not use the imager if condensation has formed. Wait until the condensation
has evaporated. You can speed up the evaporation time by moving the imager to a dryer
location.
Ventilation slots and holes are provided on the sides and rear of the imager. Place the imager on
a hard level surface and locate it at least 10 cm (4 in.) from walls to ensure proper ventilation