4. Adjustable parameters

(menu sructure: see fold out page)
Switch-on point: Upper limit value at which the output changes
its switching status.
°C
°F
Switch-off point: Lower limit value at which the output chan-
ges its switching status.
°C
°F
rP1 is always lower than SP1. The unit only accepts values which
are lower than SP1.
Changing the switch-on point also changes the switch-off point
(the hysteresis remains constant).
If the hysteresis is higher than the new switch point, it is auto-
matically reduced (rP1 is set to the minimum setting value).
Configuration of the switching output
4 switching functions can be set:
Hno = hysteresis / normally open
Hnc = hysteresis / normally closed
Fno = window function / normally open
Fnc = window function / normally closed
Lower end of analog output
Measured value for which the output signal is 4mA / 0V.
°C
°F
Upper end of analog output
Measured value for which the output signal is 20mA / 10V.
°C
°F
Minimum distance between ASP and AEP = 10°C / 18°F
All manuals and user guides at all-guides.com
setting range
-39.5 ... +125
-39 ... +257
setting range
-40 ... +124.5
-40 ... +256
setting range
-40 ... +115
-40 ... +239
setting range
-30 ... +125
-22 ... +257
in steps of
0.5
1
in steps of
0.5
1
in steps of
0.5
1
in steps of
0.5
1
17