Herunterladen Inhalt Inhalt Diese Seite drucken

Restoring The Factory Settings; Master/Slave Operation; Preparing The Devices - VOLTCRAFT LSP-1362 Bedienungsanleitung

Labornetzgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

j) Restoring the factory settings
To restore the factory settings, proceed as follows:
1. Press and hold the menu button to go to menu mode.
2. Turn the voltage knob (7) until
3. Press the voltage knob (7) to open the factory settings menu.
on the display.
4. Turn the current knob (8) until
5. Press the current knob (8) to restore the device to the factory settings.

10. Master/slave operation

In order to achieve higher output currents, up to 30 LSP-1362 laboratory power supplies can
be connected in parallel via master/slave control. In that case, all laboratory power supplies are
controlled via the master device. The current applied to the consumer is the sum of the currents
provided by the individual laboratory power supplies. The voltage is the same throughout the
circuit and corresponds to the setpoint set on the 'master' device.
a) Preparing the devices
To prepare the devices for master/slave operation, each device must be assigned its own ID:
Master device: ID=0
1. Slave device: ID=1
2. Slave device: ID=2; etc.
To assign an ID to a device, proceed as follows:
1. Press and hold the menu button to go to menu mode.
2. Turn the voltage knob (7) until
3. Press the voltage knob (7) to open the address menu.
4. Turn the current knob (8) until the desired address ID appears on the display.
5. Press the voltage knob (7) to confirm the selected address ID.
6. Press the menu button (5) to exit menu mode.
appears on the display.
appears on the display.
appears on the display.
appears
45

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

2247646

Inhaltsverzeichnis