Herunterladen Inhalt Inhalt Diese Seite drucken

Activating/Deactivating Button Lock; Using The Sense Function - VOLTCRAFT LSP-1362 Bedienungsanleitung

Labornetzgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

h) Activating/deactivating button lock
Button lock locks the buttons and knobs on the control panel in order to prevent unintentional
entries. To activate and deactivate button lock, proceed as follows:
1. Press the button lock button (4) to activate button lock. The 'LOCK' LED (15) lights up.
2. Press the button lock button (4) again to deactivate button lock. The 'LOCK' LED (15) goes
out.
i) Using the sense function
The sense function is an automatic voltage control for the connection sockets (9, 10). For this,
two separate measuring cables are connected parallel to the connecting cables. The voltage
drop which occurs on the connection cables is measured on these two measuring cables. The
laboratory power supply automatically compensates for this voltage drop so that the set voltage
is supplied to the consumer.
1. Switch the laboratory power supply and any connected consumers off.
2. Always connect the supply cables to the output on the laboratory power supply (9, 10) to the
consumer first. Ensure that you insert them with the correct polarity.
3. Press the clamp release on the rear RS+ and RS- (18) connection sockets inwards with
a small screwdriver and insert the cables into the terminal openings. Check that they are
firmly seated.
4. Now, connect the two sense cables to the consumer, observing the correct polarity. The
conductor cross-section for the sense cables must be at least 0.34 mm
5. Always undo the connections in reverse order (sense cables followed by connection cables).
.
2
43

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

2247646

Inhaltsverzeichnis