Herunterladen Inhalt Inhalt Diese Seite drucken

Medel AIR PLUS Gebrauchsanweisung Seite 61

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Tab 1
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Tab 2
Immunity test
Electrostatic discharge (ESD)
Electrical fast transient/burst
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE UT is the a.c. mians voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic emissions
Guidance and manufacturer's declaration - electromagnetic Immunity
IEC 60601-1-2 Test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Power supply lines: ±2 kV
100 kHz repetition frequency
line(s) to line(s): ±1 kV.
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º, 225 º, 270 º
and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 250 cycle
30 A/m
50Hz/60Hz
150KHz to 80MHz: 3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Power supply lines: ±2 kV
line(s) to line(s): ±1 kV.
100 kHz repetition frequency
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º, 225 º, 270 º
and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 250 cycle
30 A/m
50Hz/60Hz
150KHz to 80MHz: 3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
61
Compliance
Group 1
Class B
Class A
Compliance
Compliance level