Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Description - Axion STIM-PRO X9+ Bedienungsanleitung

Digitaler 4-kanal nerven- und muskelstimulator
Vorschau ausblenden Andere Handbücher für STIM-PRO X9+:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
AXION

TECHNICAL DESCRIPTION

01
02
EN
03
04
05
06
07
08
09
10
11
12
13
14
15
16
62
© Copyright Gebrauchs.info
Channel
Intensity
Pulse amplitude
Output voltage
Power source
Size
Weight
Pulse rate
Pulse width
ON time
OFF time
Ramp time
Mode
TENS Burst mode B
TENS Normal mode
N
TENS Modulation
mode M
4 channels – with adjustable intensity
Adjustable, 0 – 100 mA (in steps of 1 mA)
with 500 ohm load on each channel
Asymmetrical, Bi-phasic square pulse
0 – 50 V
4pcs 1.5V AA-batteries
13.8cm (L) x 7.8cm (B) x 2.8cm (H)
276 grams (batteries included)
Adjustable, from 1 – 150 Hz, 1 Hz/step
Adjustable, from 30 – 400 microseconds,
10 µ/step
Adjustable, from 2 – 90 seconds, 1 Sec./step
Adjustable, from 0 – 90 seconds, 1 Sec./step
Adjustable, from 1 – 8 seconds, 1 Sec./step
5 TENS modes: B (Burst) N (Normal),
M (Modulation), SD1 (40% modulation),
SD2 (70% modulation) – 3 EMS modes: C
(Constant), 5 (Synchronous), A (Alternate),
P (Program-Mode) – 37 TENS programs
(set by default)
Burst rate: Adjustable, 0.5 – 5 Hz – Pulse
Width adjustable, 30 – 400 µs – Frequen-
cy fixed = 100 Hz
The pulse rate and pulse width are adjus-
table. It generates continuous stimulation
based on the setting value
Modulation mode is a combination of pul-
se rate and pulse width modulation. The
pulse rate and width are automatically
varied in a cycle pattern. The pulse width
is decreased by 50% from its original
setting in 0.5 second, then the pulse rate
is decreased by 50% from its original
setting in 0.5 second. Total cycle time is
1second. In this mode, pulse rate (1 – 150
Hz) and pulse width (30 – 400 µs) are
fully adjustable.

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis