10.1 Error codes
The Incubator has built in fault diagnostics. If a fault occurs, this system will display an error code in
the LED display to help the service engineer rectify the problem. Please see the table below for details
of error codes and other faults. In the unlikely event that a problem does occur, note down which
code / fault you observe and contact Cole-Parmer for further assistance. If you have your own service
personnel available for repair, a comprehensive service manual is available on request.
Fault
Temperature probe PT100
reading low
Temperature probe PT100
reading high
Triac fault
Rotisserie / shaker motor
stalled (SI30H only)
Software crash

11 Warranty

Cole-Parmer Ltd. warrants this instrument to be
free from defects in material and workmanship,
when used under normal laboratory conditions,
for a period of three (3) years. In the event of a
justified claim, Cole-Parmer will replace any
defective component or replace the unit free of
charge.
This warranty does NOT apply if damage is
caused by fire, accident, misuse, neglect,
incorrect adjustment or repair, damage caused
by installation, adaptation, modification, fitting
of non-approved parts or repair by
unauthorized personnel.
Cole-Parmer Ltd.
Beacon Road,
Stone, Staffordshire,
ST15 0SA, United Kingdom
Tel: +44 (0)1785 812121
Email: cpservice@coleparmer.com
Web: www.stuart-equipment.com
Safety system in operation
Automatic over-temp system
activated
Automatic over-temp system
activated
Secondary relay control activated
Cuts power to motor
Cuts mains power

12 Specification

Overall dimensions
Height:
Depth:
Width:
Incubator dimensions
Height:
Depth:
Width:
Weight:
Capacity (nominal):
Temperature range:
Temperature precision:
Temperature fluctuation: ±0.1 °C (@37°C)
Power rating:
Temperature variation:
Fault displayed by
LED displays Err and 0.1
alternatively
LED displays Err and 0.2
alternatively
Top left dot flashes in display
No movement seen but
Rotisserie red indicator light
flashes
No temperature or motor
power, all LED's blank
(model SI19 & SI30H)
435 mm
380 mm
380 mm
200 mm
230 mm
250 mm
24 Kg
18 litres
Ambient +5°C – 80 °C
±0.5 °C
350W (SI19 - 300W)
<0.25°C
5