When not in use, the pads should be placed onto the plastic film liner and
sealed in the bag.
When the pads initially lose their adhesive
quality, it is possible to reactivate their
adhesiveness by applying a fine mist of water.
Replace pads when they lose their stickiness.
Note:
Poor connection may cause discomfort and skin irritation.
Storage life of an unopened pack of electrodes is 2 years. This may
be affected by very high temperatures or very low humidity.
The pads provided are latex-free.

14. CLEANING

Clean the case and lead wires at least once a week either with an alcohol-
free antibacterial wipe or by wiping with a damp lint-free cloth and a
solution of mild soap and water. Wipe dry.
Do not immerse your TENS machine in water.
Do not use any other cleaning solution.
Expected Service life
The machine will often last for more than 5 years but is guaranteed
for 2 years. Damage to the leads or electrode pads is not covered by
the warranty.
Lead life depends greatly on use. Always handle the lead with care.
Pads should last 12-20 applications, depending on skin condition and
humidity.
Fully charged battery should last about 5 hours continuous use.
Battery should last approximately 400 charge cycles.
21