IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
The Sonic Stim + device is intended for use in the electromagnetic
environment specified below. The customer or the user of the Sonic Stim +
should ensure the device is used in such an environment.
Immunity Test
Conducted
RF
IEC 610004-6
Radiated RF
IEC 61000-4-3
IEC 6061
Compliance
test level
3 Vrms 150
3 Vrms
kHz - 80
MHz
10 V/m
10 V/m
80 MHz -
2.7 GHz
commercial environment
or hospital.
Electromagnetic
environment guidance
Portable and mobile RF
communications
equipment should be used
no closer to any point of the
Sonic Stim + device,
including cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
d=l.2√P
d=12√P,80MHz to 800MHz
d=2.3√P,800MHz to 2.7GHz
where P is the maximum
output power rating of the
transmitter In watts 队)
according to the transmitter
manufacturer and d Is the
recommended separation
distance in meters (m). Field
strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in
each frequency range.