Herunterladen Inhalt Inhalt Diese Seite drucken

TensCare Sonic Stim+ Bedienungsanleitung Seite 16

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 92
3)
Adjusting intensity
Press the MODE button to select the ultrasound intensity.
The intensity has three levels, Low (L), Medium (M) and
High (H), each level corresponds to a light indicator.
4)
Adjusting treatment time
Press the TIME button to cycle through the treatment
time (5, 10 and 15 minutes), as shown by the Time LED.
When the time is chosen, the system will start working.
During working time, the user can press the TIME button
to adjust the treatment time.
5)
Start treatment
Move the treatment head in a flat, slow, circular motion
over the skin surface area that is covered with a layer
of ultrasound transmission gel. Apply the head evenly
(in time) over the treatment area.
Caution: The device has a load detection
system for safety. When the treatment head
does not have good contact with the skin,
the device will stop treatment automatically. During this time, the TIME LED will
flash slowly (l Hz). The device will not restore treatment until the contact is good.
Caution: The device has a temperature protection function. When
the temperature of the treatment head exceeds 42°C the
treatment will automatically stop and the TIME LED will flash quickly
(2Hz). The device will not restore treatment until the temperature is below 40°C.
6)
Turn off the device
After the time duration has been completed, the device will automatically
revert to standby state. Once your therapy session has been completed, turn
off the device by sliding the ON/OFF switch downwards from ON to the OFF
position.
14

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für TensCare Sonic Stim+

Inhaltsverzeichnis