Heraeus Med
6
Troubleshooting
Fault
Aufhängung / Beweglichkeit
Light is heavy to move
Light moves too easily
Lamp falls/ rises
Optical system/ light mechanism
Light flickers
Light intensity too bright/ too dark
Lamp doesn't last long enough
Light field not round
Closure disc
Scratched or cracked closure disc
Sterilization
Sterilizable handle has too short a
life
Handles damaged or having cracks Ende der Nutzungsdauer erreicht
If a fault occurs repeatedly or cannot be remedied, please get in touch with your Heraeus Med contact partner.
56 351 001/C
Gebrauchsanweisung / Operating instructions HANAULUX
Possible cause
Brakes are set too tight
Brakes are set too loose
Spring force in weight
counterbalance / spring arm too
small/ too large
Kontaktprobleme
Wrong supply voltage
Wrong lamp(s)
Wrong supply voltage
Wrong lamp(s)
Lamp seating incorrect
Main lamp defective
Incorrect cleaning agent used,
Cleaning or disinfecting when hot
Wrong sterilization procedure
Remedy
Set brakes
Set brakes
Set spring force
Service
Service
Service
Check lamp and if necessary
replace it
Check lamp seating
Use different cleaning agent
Check sterilization procedure
Replace handles
®
blue 130/100/90
Reference
Sect. 5.3
Sect. 5.3
Sect. 5.2
---
---
---
Sect/ 5.1
Sect. 3.1
Sect. 3.2
Sect. 3.2