Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Test
a)
Band
frequency
(MHz)
(MHz)
385
380 –390
450
430 – 470
710
745
704 – 787
780
810
870
800 – 960
930
1 720
1 700 –
1 845
1 990
1 970
2 400 –
2 450
2 570
5 240
5 100 –
5 500
5 800
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or
ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
Manufacturer's declaration-electromagnetic immunity
a)
Service
Modulation
Pulse
TETRA 400
modulation
18 Hz
FM
GMRS 460,
±5 kHz
FRS 460
deviation
1 kHz sine
Pulse
LTE Band 13,
modulation
17
217 Hz
GSM 800/900,
TETRA 800,
Pulse
iDEN 820,
modulation
CDMA 850, LTE
18 Hz
Band 5
GSM 1800;
CDMA 1900;
Pulse
GSM 1900;
modulation
DECT; LTE
217 Hz
Band 1, 3, 4,
25; UMTS
Bluetooth,
WLAN, 802.11
Pulse
b/g/n, RFID
modulation
2450,
217 Hz
LTE Band 7
Pulse
WLAN 802.11 a/
modulation
n
217 Hz
Maximum
b)
power
(W)
b)
1,8
c)
2
b)
0,2
b)
2
b)
2
b)
2
b)
0,2
EN-19
IMMUNITY
Distance
TEST LEVEL
(m)
(V/m)
0,3
27
0,3
28
0,3
9
0,3
28
0,3
28
0,3
28
0,3
9
Compliance
LEVEL
(V/m)
27
28
9
28
28
28
9