4) Reasonably foreseeable improper usage, malfunctions, and hazardous failure modes.
Issue - Failure mode
dose too low leading to reduced
Irradiance too low on treated
surface
No dose at all - thus no
disinfection function
Unintended human exposure to
UV-C light on when people
UVC, sunburn-like reaction to
present in the room
the skin and serious damage to
Lamp is broken during
replacement
contaminate surface and room
Material degradation, painting
Precious goods damage
Potential Effects of
Cause(s)/Mechanism(s) of
Failure
Tambient temperature of application
shadowed surface
dust on surface
disinfection rate which may
dust on UVC source
cause pathogens remain on
surface
Longer distance between lamp and
disinfected surface
lamp replaced by different (not
Philips) lamp
luminaire is not connected
Luminaire is not on
luminaire is not switched on
driver not working
End of life or failure of driver
Lamp not working
End of life or failure of lamp
Containment safeguard not installed
Containment safeguard - check people
presence failed - Luminaire started
while people present in the room
Containment Safeguard – After
checking room if nobody is present,
the cornea
person fails to lock the room. Room
remains unlocked.
Containment safeguard - detection
sensor/presence sensor failed
Activation by non-authorized person
Luminaire is on while relamping
Hg from the lamp may
Accidental fall of lamp while
relamping
Precious goods not secured while
fade away due to UVC
lamp operating
Potential
Failure
Temp monitoring,
Keep temperature within limits described in product specification
Make sure treated surface will not be shadowed by other objects
Make sure treated surface is dust free
Make sure lamp (UVC source) is clean and free of dust
Make sure distance and irradiance level are fit correctly for a
disinfection purpose. Irradiance decreases with square of distance
(e.g. 2x longer distance results in 4x lower irradiance)
Make sure lamp is replaced by Philips and with frequency in
accordance to lamp specification
Connect luminaire according to manual instruction
Switch on luminaire with all safety measure described in user
manual
Replace driver
Replace lamp, working lamp emits bluish light. Checking must be
done in safe manner , keeping all safety precautions
Check training materials and user manual. Apply minimum
containment safety measures. Never operate luminaire without
containment safety measures installed and checked they're
working
Check training materials and user manual. Apply minimum
containment safety measures. Make sure check of people
presence is secured and effective (e.g. motion and presence
detector)
Check training materials and user manual. Apply minimum
containment safety measures. Make sure locking the room before
starting luminaire is secured and effective.
Make good commissioning of your sensor(s). Make sure other
levels of containment safety are implemented effectively (e.g.
door switch shall shut off light when doors opened)
Secure activation of luminaires can only be done by authorized
personnel (e.g. use key lock switch)
Always check and secure luminaires are off before entering room.
Be careful and focused while relamping. Follow replacement
instruction. In case of breakage follow instruction how to clean Hg
contamination.
Make sure all precious goods are removed or hidden/protected
from direct UVC application