To reduce the risk associated with hazardous voltage:
• Use Indoors only.
• Only use device with power supply part number 26-1022-0552-8 supplied
with the 3M Petrifilm Plate Reader Advanced.
• Do not spill liquids on device or immerse in liquids.
• Always unplug the 3M Petrifilm Plate Reader Advanced before cleaning the
device.
• The Personal Computer (PC) that the 3M Petrifilm Plate Reader Advanced
connects to must be an approved certified device (mark with UL, CSA, ETL,
TUV, or other required regulatory safety agency in your region) and meet
requirements specified in the user manual.
• The USB cable must only be connected to an approved Listed/certified
device.
• The external power supply serves as the mains supply to disconnect the
device. Do not position the equipment so that it is difficult to disconnect the
device.
• The external power supply is a class I device. Only plug into an appropriate
AC grounded outlet.
To reduce the risk associated with EMC interference:
• Modifications to this device or power supply shall not be made without the
written consent of 3M, Company. Unauthorized modifications may void
the authority granted under Federal communication Rules permitting the
operation of this device.
To reduce the risk associated with misinterpretation of results:
• Users should be trained Microbiologists that are able to manually interpret
3M Petrifilm Plates and have been trained and understand fully how to
operate the 3M Petrifilm Plate Reader Advanced and the 3M Petrifilm Plate
Manager.
• Users should insert 3M Petrifilm Plates into the 3M Petrifilm Plate Reader
Advanced in the correct orientation.
• Users should not feed 3M Petrifilm Plates through the device that have
greater than 25% Liquefier colonies.
• Follow all instructions provided in the 3M Petrifilm Plate Reader Advanced
User Manual.
• 3M has not documented 3M Petrifilm Plates for use in industries other
than food or beverage. For example, 3M has not documented 3M Petrifilm
Plates for testing water, pharmaceuticals, or cosmetics. 3M has not tested or
documented 3M Petrifilm Plates for the use in all food matrices.
• Do not use any damaged or third-party parts including the user manual and
important safety information with the 3M Petrifilm Plate Reader Advanced.
• All parts must be serviced by an authorized 3M Service Center.
W WARNING
CAUTION