3.4 Entering Power Coefficients

For accurate control of the converter output power, the R&S ZVA analyzer must know
the (non-linear) current-power characteristic of the converters. The characteristic is suf-
ficiently described by a third-order polynomial. A label with the four polynomial coeffi-
cients c
The power coefficients must be adjusted whenever a new converter is connected to an
analyzer port. To enter the coefficients, proceed as follows:
1. Activate the frequency converter mode following the first steps in
"Activating the Frequency Converter
2. In the "Frequency Converter" dialog, press "Coefficients".
3. In the "Power Coefficients" dialog opened, clear "Use default coefficients" and
enter your coefficients into the appropriate table row.
The numbers of the table rows denote the analyzer ports for the converters.
4. Repeat the last step for all frequency converters in the test setup.
Getting Started 1307.7197.62 ─ 08
, c
, c
, and c
is affixed to each converter.
0
1
2
3
Entering Power Coefficients
Mode",
on page 20.
Basic Operation
Chapter 3.3,
21