15 . 1 .2 . Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Zen-O lite™ is intended for use in the electromagnetic environment specified below. The
customer or the user of the concentrator should assure that it is used in such an environ-
ment.
IMMUNITY
TEST
Electrostatic
EN
discharge
(ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short inter-
ruptions,
and voltage
variations
on power
supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz) ma-
gnetic field
IEC 61000-4-8
IEC 60601 TEST
COMPLIANCE
LEVEL
± 2, 4, 6, 8kV
± 2, 4, 6, 8kV
contact
contact
± 2, 4, 8, 10, 15kV
± 2, 4, 8, 10, 15kV
air
air
± 2kV for power
± 2kV for power
supply lines
supply lines
± 1kV for input/
± 1kV for input/
output lines
output lines
± 1kV differential
± 1kV differential
mode
mode
± 2kV common
± 2kV common
mode
mode
< 5% UT
< 5% UT
(>95% dip in UT)
(>95% dip in UT)
for 0.5 cycle
for 0.5 cycle
40% UT
40% UT
(60% dip in UT)
(60% dip in UT)
for 5 cycles
for 5 cycles
70% UT
70% UT
(30% dip in UT)
(30% dip in UT)
for 25 cycles
for 25 cycles
<5% UT
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 5 sec
for 5 sec
30 A/m
30 A/m
ELECTROMAGNETIC
LEVEL
ENVIRONMENT/GUIDANCE
Floors should be wood, concrete, or
ceramic tile.
If floors are covered with synthetic
material, the relative humidity should
be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should
be that of a typical commercial or
hospital environment. If the user
of the Zen-O  lite™ Portable Oxygen
Concentrator required continued
operation during power main inter-
ruptions, it is recommended that
the concentrator be powered from
an uninterruptible power supply or
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or hospital
environment.
36/340