REF. DESCRIPTION
1
Downwards arrow When pressed, it decreases the value of the
2
Upwards arrow
3
V
I
4
5
Set 2
Set 1
6
7
Amp
8
KV
Display
9
10
3Kv / 5Kv / 7Kv
11
PROTECT
HV ON/OFF
12
13
HV
14
REMOTE
15
START
PROT
16
MENU
Downloaded from
www.Manualslib.com
with (12) set to OFF
function selected by means of the (16) pushbutton.
When pressed, it increases the value of the
function selected by means of the (16) pushbutton.
Control disabled.
Control disabled.
When it lights up the display (9) shows the Set
Point 2 value when selected by means of the (16)
pushbutton.
When it lights up the display (9) shows the Set
Point 1 value when selected by means of the (16)
pushbutton.
It shows the voltage or current value selected by
means of the pushbutton (16).
The LED on corresponds to the maximum controller
output voltage.
When it lights up the "Protect" operating mode is
activated.
When it flashes the controller is in setting mode.
It switches on the high voltage.
When it lights up the MidiVac is controlled by an
host computer.
When pressed it activates the "Protect" operating
mode.
LED (11) flashes during the setting operations.
If the MidiVac is in "Remote" mode, when pressed
it switches the operating mode to "Local".
If the MidiVac is in "Local" mode, when pressed it
allows to access to the setting functions.
manuals search engine
INSTRUCTIONS FOR USE
DESCRIPTION
with (12) set to ON
When pressed it changes the maximum output
voltage from 7 kV to 5 kV and from 5 kV to 3 kV.
When pressed it changes the maximum output
voltage from 3 kV to 5 kV and from 5 kV to 7 kV.
When pressed the display (9) shows the value of the
output voltage to the pump.
When pressed the display (9) shows the pump
absorbed current.
When it lights up the current value has reached the
Set Point 2 value.
When it lights up the current value has reached the
Set Point 1 value.
When it lights up the display (9) shows the pump
absorbed current.
When it lights up the display (9) shows controller
output voltage.
It shows the controller output voltage, or the pump
absorbed current.
The LED on corresponds to the maximum controller
output voltage.
When it lights up the "Protect" operating mode is
activated.
It switched off the high voltage.
When it lights up the high voltage is present.
When it lights up the MidiVac is controlled by an host
computer.
When pressed it activates the "Protect" operating
mode.
If the MidiVac is in "Remote" mode, when pressed it
switches the operating mode to "Local".
If the MidiVac is in "Local" mode, when pressed it
allows to switch to "Remote".
22
87-900-080-01 (D)