Verfügbare Sprachen

Instruction Manual CertoClav Vacuum Pro
Electromagnetic immunity
The Steam sterilizer is intended for use in the electromagnetic environment specified below. The customer or the user of the Steam sterilizer
should assure that it is used in such an environment.
Immunity
test
IEC 60601 test level
Conducted
3 Vrms
RF
150 kHz to 80 MHz
IEC 61000-
4-6
3 V/m
80 MHz to 2,5 GHz
Radiated RF
IEC 61000-
4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 these guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in
which the Steam sterilizer is used exceeds the applicable RF compliance level above, the Steam sterilizer should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Steam sterilizer.Over
the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
Jan. 2018
Compliance level
Electromagnetic environment - guidance
3 Vrms
Portable and mobile RF communications
equipment should be used no closer to any
150 kHz to 80 MHz
part of the Steam sterilizer including cables,
3 Vrms
than the recommended separation distance
calculated from the equation applicable to the
3 V/m
frequency of the transmitter.
80 MHz to 2,5 GHz
3 V/m
d = 1,2 √ ��
d = 1,2 √ �� 80 MHz to 800 MHz
d = 2,3 √ �� 800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey, a should be less than the compliance
level in each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
57