Instruction Manual CertoClav Vacuum Pro
Electromagnetic immunity
The Steam sterilizer is intended for use in the electromagnetic environment specified below. The customer or the user of the Steam sterilizer
should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601
test level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/output
lines
1 kV line(s) to line(s)
2 kV line(s) to earth
U
<5 %
T
U
(>95 % dip in
T)
for 0,5 cycle
U
40 %
T
U
(60 % dip in
T)
for 5 cycles
U
70 %
T
U
(30 % dip in
T)
for 25 cycles
U
<5 %
T
U
(>95 % dip in
T)
for 5 sec
3 A/m
Jan. 2018
Compliance level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for input/output
lines
1 kV line(s) to line(s)
2 kV line(s) to earth
U
<5 %
T
U
(>95 % dip in
T)
for 0,5 cycle
U
40 %
T
U
(60 % dip in
T)
for 5 cycles
U
70 %
T
U
(30 % dip in
T)
for 25 cycles
U
<5 %
T
U
(>95 % dip in
T)
for 5 sec
3 A/m
56
Version 7
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
If the
user
Steam sterilizer requires continued
operation during power
mains
interruptions,
recommended
that the
sterilizer be powered from an
uninterruptible
power supply
battery.
Power
frequency magnetic
should be at levels characteristic of
a typical location
in a
commercial
or
environment.
of the
it
is
Steam
or a
fields
typical
hospital