Herunterladen Inhalt Inhalt Diese Seite drucken

CertoClav Vacuum Pro Serie Bedienungsanleitung Seite 110

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Instruction Manual CertoClav Vacuum Pro
E9
In
the
sterilization
period,
sterilization
pressure
is 0.3 bar.
E10
The electronic locker is
in the wrong condition.
E11
The electronic locker is
in the wrong condition
E12
The vacuum has not
reached -70Kpa during
at least
3 vacuum
phases.
E99
Communication
between
the
display
and
mainboard
working.
No water inside the tank. Heater
inside steam generator is broken.
Electronic
locked
Circuit board damaged.
Electronic
locked
Circuit board damaged.
Vacuum pump is damaged, fan is
damaged.
Cable is not connected. The CPU
front
IC has loosened.
the
isn't
Jan. 2018
Check heaters, water pump, water
level. Water pump is damaged.
damaged.
Check the electronic locker and
circuit board.
damaged.
Check the electronic locker and
circuit board.
Check the vacuum pump and the fan.
Check the cable between the front
display and the mainboard.
Check if the microcontroller IC at the
mainboard is not loose.
49