Herunterladen Inhalt Inhalt Diese Seite drucken

GE Druck PV 210 Handbuch Seite 7

Niederdruck-handpumpe
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 13
1. Introduction
The PV 210 is a portable source of pressure and vacuum. Each pump includes a pressure/
vacuum selector, a volume control for fine adjustment and an adjustable pressure relief valve to
prevent damage to sensitive instruments.
2. Operation
Note: Wherever possible, use o-ring seals in the BSP connection ports this is the
recommended method of sealing.
Key to Figure #1
1.
2.
3.
4.
5.
6.
7.
8.
2.1 Hose/Adaptors (Figure #1)
To attach a hose (5) and adapter (6) to a connector (4), turn the knurled nut on the connector
counterclockwise.
© 2005 Baker Hughes, a GE company – All rights reserved.
Downloaded from
www.Manualslib.com
WARNING
Before applying pressure, make sure all connections are
correct and equipment is internally clean and free from damage.
Make sure that all equipment is to the correct pressure rating.
Do not exceed the maximum operating pressure stated in the specification.
Observe the relevant health and safety precautions.
CAUTION
If PTFE tape is used to seal NPT threads, ensure that only a
sufficient amount is used to achieve pressure seal. If excess tape is used,
particles can become loose during the connector mating process and enter
the pump, potentially leading to loss of pump performance or pressure leaks.
INFORMATION
use the PV 210 to pressurize small volumes.
Pressure release valve.
Fine adjust vernier.
Pressure/Vacuum selector.
Quick-fit connectors.
1 m (3.3 ft) x 3 mm (1/8") flexible nylon hose (x2).
1/4" female BSP or NPT adaptor.
Pump handle.
Adjustable pressure relief valve.
manuals search engine
Because the PV 210 only has a small displacement, only
PV 210 Instruction Manual-English | 3

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis