Herunterladen Diese Seite drucken

Delta OHM HD 9016 Handbuch Seite 4

Mikroprozessgesteuerte thermometer mit digitalanzeige
Vorschau ausblenden Andere Handbücher für HD 9016:

Werbung

DATI TECNICI
Ingressi: 2.
CAMPO DI MISURA STRUMENTO:
Termocoppia K: ( -200...+1370°C) ( -328...+1999°F)
Termocoppia J: ( -100...+ 750°C) ( -148...+1382°F)
Termocoppia T: ( -200...+ 350°C) ( -328...+ 662°F)
Termocoppia E: ( -200...+ 750°C) ( -328...+1382°F)
Termocoppia R: (+200...+1480°C) (+392...+1999°F)
Termocoppia S: (+200...+1480°C) (+392...+1999°F)
Termocoppia B: (+200...+1800°C) (+392...+1999°F)
Collegamento sonde: 2 prese miniatura per termocoppia K,
giunto freddo compensato.
Risoluzione: ≤ 199,9°C (+199,9°F) 0,1°C (0,1°F)
≥200°C (+200°F) 1°C (1°F)
Precisione strumento: Da 0...+199,9°C (+199,9°F) ±0,1% della
lettura ±0,4°C (±0,7°F) ±1 digit. Da 200°C (200°F) a fondo scala o
da -0,1°C (31,8°F) a -200°C (-328°F) ±0,2% della lettura +1°C
(+1,8°F) ±1 digit.
Questa precisione è valida per temperatura ambiente 25°C ±5°C.
Influenza della temperatura nel campo di lavoro dello
strumento -5...+18°C o +25...+50°C:
-200... -50°C: 0,4°C +0,01°C/°C
-50... +200°C: 0,2°C +0,01°C/°C
+200...+800°C:
3°C
Frequenza di conversione: 1 secondo.
Temperatura di lavoro strumento: 0...50°C.
Umidità relativa: 0...90% U.R.
Funzioni strumento: Cambio scala automatico, HOLD,
spegnimento automatico, memorizzazione del valore di
temperatura massimo, del valore di temperatura minimo, misure
relative, differenza fra gli ingressi, selezione del tipo di
termocoppie.
Display: LCD 3
1
/
digit altezza cifre 12,5 mm, simboli HOLD,
2
RCD, REL, MAX, MIN, °C, °F, A, B, A-B.
Alimentazione: Batteria 9V IEC 6LF22.
Durata batteria: indicativo 100 ore, batterie alcaline.
Peso strumento: 280 gr.
Dimensioni: 215x73x38 mm.
CODICE DI ORDINAZIONE:
HD 9016 kit: Strumento HD 9016, batteria 9V, manuale
d'istruzione, valigetta. Dimensioni: 450x300x105 mm,
peso 1800 gr.
TECHNISCHE DATEN
Eingänge: 2.
MEßBEREICH DES GERÄTES:
Thermoelement K: ( -200...+1370°C) ( -328...+1999°F)
Thermoelement J: ( -100...+ 750°C) ( -148...+1382°F)
Thermoelement T: ( -200...+ 350°C) ( -328...+ 662°F)
Thermoelement E: ( -200...+ 750°C) ( -328...+1382°F)
Thermoelement R: (+200...+1480°C) (+392...+1999°F)
Thermoelement S: (+200...+1480°C) (+392...+1999°F)
Thermoelement B: (+200...+1800°C) (+392...+1999°F)
Sondenanschluß: 2 Miniatursteckbuchsen für Thermoelemente K, kaltgeschweißt, mit Kompensation.
Auflösung: ≤±199,9°C (+199,9°F) 0,1°C (0,1°F)
Anzeigegenauigkeit: Von 0...+199,9°C (+199,9°F) ±0,1% der Anzeige ±0,4°C (±0,7°F) ±1 digit. Von
200°C (200°F) bis Skalenendwert oder von -0,1°C (31,8°F) bis -200°C (-328°F) ±0,2% der Anzeige
+1°C (+1,8°F) ±1 digit. Diese Genauigkeit gilt bei einer Umgebungstemperatur von 25°C ±5°C.
Einfluß der Umgebungstemperatur -5...+18°C oder +25...+50°C:
-200... -50°C: 0,4°C +0,01°C/°C
-50... +200°C: 0,2°C +0,01°C/°C
+200...+800°C:
3°C
Wandlungsrate: 1 Messung pro Sekunde.
Arbeitstemperatur des Gerätes: 0...50°C.
Relative Luftfeuchtigkeit: 0...90% r.F.
Funktionen des Gerätes: Automatischer Skalenwechsel, HOLD, automatische Abschaltung,
Speicherung des maximalen und minimalen Temperaturwertes, Relativmessungen,
Differenzmessungen der Eingänge, Anwahl des Thermoelementetyps.
1
Display: LCD 3
/
digit Zahlenhöhe 12,5 mm, Symbole HOLD, RCD, REL, MAX, MIN, °C, °F, A, B, A-B.
2
Stromversorgung: Batterie 9V IEC 6LF22.
Lebensdauer der Batterie: ca. 100 Stunden bei Alkalibatterien.
Gewicht des Gerätes: 280 g.
Abmessungen: 215x73x38 mm.
BESTELL-NR:
HD 9016 kit: Gerät HD 9016, 9V-Batterie, Gebrauchsanweisung, Koffer. Abmessungen 450x300x105
mm. Gewicht 1800 g.
SIT CALIBRATION
CENTRE N° 124
TECHNICAL DATA
Inputs: 2.
INSTRUMENT RANGE:
Thermocouple K: ( -200...+1370°C) ( -328...+1999°F)
Thermocouple J: ( -100...+ 750°C) ( -148...+1382°F)
Thermocouple T: ( -200...+ 350°C) ( -328...+ 662°F)
Thermocouple E: ( -200...+ 750°C) ( -328...+1382°F)
Thermocouple R: (+200...+1480°C) (+392...+1999°F)
Thermocouple S: (+200...+1480°C) (+392...+1999°F)
Thermocouple B: (+200...+1800°C) (+392...+1999°F)
Probe connections: 2 miniature thermocouple K sockets,
compensated cold junction.
Resolution: ≤ +199.9°C (+199.9°F) 0.1°C (0.1°F)
≥ +200°C (+200°F) 1°C (1°F)
Instrument precision: From 0°C to +199.9°C (+199.9°F) ±0.1%
of reading ±0.4°C (±0.7°F) ±1 digit. From 200°C (200°F) to full
scale or from -0.1°C (31.8°F) to -200°C (-328°F) ±0.2% of reading
+1°C (+1.8°F) ±1 digit.
This precision applies to an environment temperature 25°C ±5°C.
Influence of temperature in the instrument working range
-5...+18°C o +25...+50°C:
-200... -50°C: 0.4°C +0.01°C/°C
-50... +200°C: 0.2°C +0.01°C/°C
+200...+800°C:
3°C
Conversion frequency: 1 second.
Instrument working temperature: 0...50°C.
Relative humidity: 0...90% R.H.
Instrument functions: Automatic change of scale, HOLD,
automatic cut-out, MAX/MIN storage, relative measurements,
difference between inputs, selection of thermocouple type.
Display: LCD 3
1
/
digit figure height 12.5 mm, symbols HOLD,
2
RCD, REL, MAX, MIN, °C, °F, A, B, A-B.
Power supply: 9V battery IEC 6LF22.
Battery life: approximately 100 hours, alkaline battery.
Instrument weight: 280 gr.
Dimensions: 215x73x38 mm.
ORDER CODES:
HD 9016 kit: Instrument HD 9016, 9V battery, instructions
manual, diplomatic carrying case. Dimensions 450x300x105 mm,
weight 1800 gr.
≥±200°C (+200°F) 1°C (1°F)
DELTA OHM SRL - VIA G. MARCONI, 5
35030 CASELLE DI SELVAZZANO (PD) - ITALY
TEL. 0039-0498977150 r.a. - FAX 0039-049635596
e-mail: deltaohm@tin.it - Web Site: www.deltaohm.com
CARACTERISTIQUES TECHNIQUES
Entrées: 2.
ETENDUE DE MESURE:
Thermocouple K: ( -200...+1370°C) ( -328...+1999°F)
Thermocouple J: ( -100...+ 750°C) ( -148...+1382°F)
Thermocouple T: ( -200...+ 350°C) ( -328...+ 662°F)
Thermocouple E: ( -200...+ 750°C) ( -328...+1382°F)
Thermocouple R: (+200...+1480°C) (+392...+1999°F)
Thermocouple S: (+200...+1480°C) (+392...+1999°F)
Thermocouple B: (+200...+1800°C) (+392...+1999°F)
Raccordement sondes: 2 prises miniature pour thermocouple K,
joint froid compensé.
Résolution: ≤ 199,9°C (+199,9°F) 0,1°C (0,1°F)
Précision: De 0...+199,9°C (+199,9°F) ±0,1% de la lecture
±0,4°C (±0,7°F) ±1 digit. De 200°C (200°F) pleine échelle ou de
-0,1°C (31,8°F) à -200°C (-328°F) ±0,2% de la lecture +1°C
(+1,8°F) ±1 digit.
Cette précision est valable pour température ambiante 25°C ±5°C.
Influence de la température dans la courbe d'utilisation de
l'appareil -5...+18°C ou +25...+50°C:
-200... -50°C: 0,4°C +0,01°C/°C
-50... +200°C: 0,2°C +0,01°C/°C
+200...+800°C:
Temps d'intégration: 1 s.
Domaine d'utilisation: 0...50°C.
Humidité relative: 0...90% H.R.
Fonctions: sélection de gamme automatique, [HOLD], arrêt
automatique, mémorisation de la valeur de température maximum
et de la valeur de température minimum, mesures relatives,
différence entre les entrées, sélection du type de couple
thermoélectrique.
Afficheur: LCD 3
touches [HOLD], [RCD], [REL], [MAX], [MIN], °C, °F, A, B, A-B.
Alimentation: Pile 9V IEC 6LF22.
Autonomie: approximativement 100 h, (pile alcaline).
Masse: 280 g.
Dimensions: 215x73x38 mm.
CODE DE COMMANDE:
HD 9016 kit: Instrument HD 9016, pile 9V, manuel d'utilisations,
mallette. Dimensions 450x300x105 mm, masse 1800 g.
DATOS TECNICOS
Ingresos: 2.
RANGO DE MEDICIO ´N DEL INSTRUMENTO:
Termopar K: ( -200...+1370°C) ( -328...+1999°F)
Termopar J: ( -100...+ 750°C) ( -148...+1382°F)
Termopar T: ( -200...+ 350°C) ( -328...+ 662°F)
Termopar E: ( -200...+ 750°C) ( -328...+1382°F)
Termopar R: (+200...+1480°C) (+392...+1999°F)
Termopar S: (+200...+1480°C) (+392...+1999°F)
Termopar B: (+200...+1800°C) (+392...+1999°F)
Connección sondas: 2 connectores miniatura para termopar K, juntura fría compensada.
Resolución: ≤199,9°C (+199,9°F) 0,1°C (0,1°F)
Precisión instrumento: De 0...+199,9°C (+199,9°F) ±0,1% de la lectura ±0,4°C (±0,7°F) ±1 digit. De
200°C (200°F) a fondo escala o de -0,1°C (31,8°F) a -200°C (-328°F) ±0,2% de la lectura +1°C
(+1,8°F) ±1 digit. Esta precisión es valida para temperatura ambiente 25°C ±5°C.
Influencia de la temperatura en el campo de trabajo del instrumento -5...+18°C ó +25...+50°C:
-200... -50°C: 0,4°C +0,01°C/°C
-50... +200°C: 0,2°C +0,01°C/°C
+200...+800°C:
3°C
Frequencia de conversión: 1 segundo.
Temperatura de trabajo del instrumento: 0...50°C.
Humedad relativa: 0...90% H.R.
Funciones del instrumento: Cambio de escala automático, HOLD, apagado automático,
memorización del valor de temperatura máximo, del valor de temperatura mínimo, mediciones
relativas, diferencia entre los ingresos, selección del tipo de termopar.
1
Display: LCD 3
/
dígitos altura de las cifras 12,5 mm, símbolos HOLD, RCD, REL, MAX, MIN, °C,
2
°F, A, B, A-B.
Alimentación: Pila 9V IEC 6LF22.
Duración pila: indicativo 100 horas, pila alcalina.
Peso instrumento: 280 gr.
Dimensiones: 215x73x38 mm.
CODIGO DE PEDIDO:
HD 9016 kit: Instrumento HD 9016, pila 9V, manual de instrucciones, maletín.
Dimensiones 450x300x105 mm. Peso 1800 gr.
Miglioriamo in continuazione i nostri prodotti, ci riserviamo il diritto di apportare modifiche senza preavviso.
We improve continually our products and reserve us the right to modify them without prior notice.
Nous améliorons continuellement nos produits, nous réservons le droit de le modifier sans préavis.
Wir entwickeln unsere Produkte weiter und behalten uns das Recht der Änderung vor.
Mejoramos continuamente nuestros productos, nos reservamos el derecho de modificarlos sin previo aviso.
≥ 200°C (+200°F) 1°C (1°F)
3°C
1
/
digits hauteur chiffres 12,5 mm, symboles
2
≥200°C (+200°F) 1°C (1°F)
SAFIGRAF 03/02

Werbung

loading