Verfügbare Sprachen

Calibrate the scale
Only calibrate this scale if you are familiar with the technic and have the right aproved massweights
1
Switch off the scale
2
Press UNIT+ON/OFF keys together
Display shows the AD value
3
Press ON/OFF key
4
Put the displayed weight on the platter, confirm with On/Off key
5
PASS will appear and scale goes to weighingmode
Use tare function
Put the scale on a stable surface
Switch on the scale and wait for zero
put an empty container on the platter and press ZERO, display shows zero again
Your next weight will be NET weight
to erease tare weight just empty the platter and press ZERO again