Configuring the pulse output
Press the key »Param« repeatedly until »
displayed light on dark background.
Select the desired signal using
triggered mandatory stroke
or
mandatory stroke
or
alarm
Default setting:
mandatory stroke
Configuring RS 232 interface
To select data transmission rate (baudrate) and parity check
mode.
Press the key »Param« repeatedly until »Baudrate« is
displayed light on dark background.
Select the baudrate with the keys
9600
or
2400
or
1200
Press the key »Param« to select »Parity«.
Select parity check mode with the keys
NONE
or
EVEN
or
ODD
When using the printer select »NONE«.
Default settings:
Baudrate:
9600
Parity:
NONE
Stop bits:
1 (fixed)
Data bit:
8 (fixed)
Setting the printer
The HP ThinkJet must be configured as follows (cf. Owner's
Manual HP ThinkJet).
Baudrate:
same as Babylog 8000
Parity:
NONE
Data bits:
8
Handshake mode:
XON/XOFF
Alternate Control Sequence Mode
« is
or
:
or
:
or
: