Operation
Recording analogue data
One analogue signal each via Analog1 and Analog2.
Voltage range: 0 to 10 V
Connect recorder (input impedance ≥ 1 MΩ) with
cable 83 06 487 (cf. Technical Data, page 29).
Selecting signal and range, page 26.
Recording
– signal via the pulse output
The
–indicates certain events during ventilation, e.g.,
each mandatory stroke, or each triggered mandatory stroke.
The signal voltage has either H(igh) level or L(ow) level
(cf. Technical Data, page 29).
Depending on the setting, the pulse has the following
waveform:
Mandatory stroke
H level during mandatory strokes, otherwise L level.
Triggered mandatory stroke
H level during triggered mandatory strokes, otherwise L level.
Alarm
L level during an alarm condition, otherwise H level.
Connect recorder (input impedance ≥ 1 MΩ) via a cable
83 06 487.
Select signal, cf. configuring pulse output, page 27.
Analog1
Analog2
H
mandatory stroke
L
H
mandatory stroke
L
trigger
H
alarm condition
L
Analog1
Analog2
Operation
19