Jumper connections
When a 4-wire load cell is attached to terminal block J4, jumpers W1 and W2 must be installed.
When a 6-wire load cell is attached to terminal block J4, jumpers W1 and W2 must be removed.
For load cells with an extra ground shield wire: Connect the shield to the center position (GND)
of J4.
RS232 Interface Cable to T51XW
Pass the optional RS232 cable through the cable gland and attach it to the terminal block J7. Tighten the cable gland to
secure the cable and maintain a tight seal.
Footswitch to T51P or T51XW
Pass the optional footswitch cable through the cable gland and attach it
to the terminal block J9. Tighten the cable gland to secure the cable and maintain a tight seal.
3. OPERATION
3.1 Overview of Display and Controls
Terminal Block J4
Pin
1
2
3
4
5
6
7
Pin
Connection
1
RTS
2
TXD
3
RXD
4
CTS
5
Ground
Connection
+ Excitation
+ Sense
+ Signal
Ground
– Signal
– Sense
– Excitation
EN-3