After reprocessing

Checking the operational readiness
Prerequisites:
The product has been assembled and prepared
so that it is ready for operation.
Procedure:
1
Check the operational readiness, see chapter
"Preparation and operation".

Maintenance

Safety information

WARNING
Risk due to inappropriately reprocessed prod-
ucts
The product may be contaminated with infec-
tious agents.
Before service is performed and before the
product is sent back for repair, reprocess the
product in accordance with the chapter "Repro-
cessing".
WARNING
Risk if service is not performed regularly
Wear and material fatigue of the components
may lead to device failure and malfunctions.
Perform service at the specified intervals.
38
WARNING
Risk if service is not performed properly
Personal injury and property damage may oc-
cur if service is not performed properly.
Service must be performed by those user
groups that are assigned to the particular mea-
sure.
WARNING
Risk if maintenance is not performed properly
If the device is connected to the gas supply
during maintenance, there is a risk of personal
injury and property damage.
Before performing maintenance, disconnect all
gas connections from the gas supply.
WARNING
Risk of user injury
Do not disassemble the pressure dome of the
pressure reducer while it is under pressure.

Definition of maintenance concepts

Concept
Definition
Service
All measures (inspection, mainte-
nance, repair) intended to maintain
or restore the functional integrity of a
product

Inspection

Measures intended to determine and
assess the current state of a product
Mainte-
Regular specified measures intend-
nance
ed to maintain the functional integrity
of a product

Repair

Measures intended to restore the
functional integrity of a product after
a failure
Inspection
.
Checks
Interval
Personnel respon-
sible
Inspection
Every
Service personnel
2 years

Preventive maintenance

Compo-
Inter-
Measure
nent
val
Sealing of
Every
Replacement
pressure
2 years
reducer/
gas cylin-
der
O-ring for
Every
Replacement
NIST hose
4 years
Pressure
Every
Major over-
reducer
6 years
haul, replace-
ment of mem-
branes and
sealing ele-
ments
Repair
Dräger recommends that all repairs are performed
by DrägerService and that only authentic Dräger re-
pair parts are used.
Instructions for use OxyLine