Herunterladen Inhalt Inhalt Diese Seite drucken

Connection To The Pump - Agilent Technologies TV 902 Bedienungshandbuch

Vorschau ausblenden Andere Handbücher für TV 902:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29

Connection to the Pump

Figure 23
Also the pump cable is shielded and is equipped with ferrite (see the
following picture). Pump cable length must not exceed 1,5 m.
Figure 24
The turbopump is connected to the controller through an 8-pin
connector. Pins B, C and D are the 3-phase supply to the motor, pins
A and F are connected to the temperature sensor (NTC type, 30 KΩ
resistance at 25 °C) and pin E is connected to the pump ground; pins
G and H are not connected.
TV 902 Navigator User Manual / 87-900-999-01
Technical Information

Connection to the Pump

(A)
18
285/292

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis