Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies TV 902 Bedienungshandbuch Seite 267

Vorschau ausblenden Andere Handbücher für TV 902:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29
Figure 12
The set point output stays at 0 Vdc until the frequency becomes
higher than 505 Hz (that is 500 Hz + 1% of 500 Hz), then the output
goes at 24 Vdc and stays at 24 Vdc until the frequency becomes lower
than 495 Hz (that is 500 Hz – 1% of 500 Hz).
It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
The Navigator Software (optional) allows the oper-ator to set all the
NOTE
programmable feature.
TV 902 Navigator User Manual / 87-900-999-01
reference quantity: frequency
threshold: 582 Hz
hysteresis: 2%
activation type: high level
delay time: 0 second
Technical Information
(A)
18
Interconnections
267/292

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis