Maintenance - Baaske Medical MI 1005 Handbuch

Netzwerkisolator
Vorschau ausblenden Andere Handbücher für MI 1005:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3. Installation
The installation of the MI 1005 should be done near
the medical appliance/system. The safety advice in
Chapter -1.2 – Connection and Operation- is to be
observed!

4. Maintenance

For maintaining a safe technical condition, an annu-
al inspection is recommended, which includes the
following checks:
1. Examination for external damage
(housing, network connections, legibility
of labeling, dirt, etc.) of the MI 1005 .
2. Examination for substances which may have
penetrated, in
3. Function test.
4. Isolation test. The isolation resistance of
the MI 1005 should not exceed 20 MOhm.
For testing, at both connections all
data cables as well as the screen should
always be shut down briefly. The isolation
resistance should be measured between
signal entry and signal exit.
5. Availability and sufficiency of the docments.
Page 26
MI 1005 Manual V3
11-2019