Calibrating the temperature sensor
5.
Thermometer deviation is set to 0 ° C from
production. The values indicated by the
thermometer can be adjusted when the
values displayed on the incubator display
differ from the values given in the calibrated
thermometer.
5.1. Press and hold SET button for 3s.
5.2. Press SET again and find „CA"
5.3. With " ↑ " and " ↓ " find desired deviation
5.4. Press SET button to save and leave
NOTE: The deviation value may also be negative
if the value of incubator is higher than the value
of calibrated thermometer. The set value is the
difference between the values indicated by the
incubator and the values given by the calibrated
thermometer.