Herunterladen Inhalt Inhalt Diese Seite drucken

H&O Equipments CryoPen M Gebrauchsanweisung Seite 12

Vorschau ausblenden Andere Handbücher für CryoPen M:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Step 1
I
Empty the loading cavity
Before loading a new cartridge, release any unused gas from the current
cartridge by unscrewing the applicator followed by pressing the button.
Then unscrew the device. Discard the used cartridge.
Regarding CryoPen X+/CryoProbe X+, CryoPen O/CryoProbe O,
CryoPen XP/CryoProbe XP be careful not to throw the converter 16/8g.
Please ensure that the used filter does not remain inside the cavity.
Step 2
I
Load a new cartridge
The cartridges are packed in peel-pack blisters with an assembled filter
and cartridge opener included. The filters are used for protection
of the applicator, the opener for puncturing the cartridge.
CryoPen M/CryoProbe M can be used with 8g cartridges only.
CryoPen X+/CryoProbe X+, CryoPen O/CryoProbe O, CryoPen XP/CryoProbe XP
can be used with 8g or 16g cartridges. Ensure that the loading cavity is empty
of all residue. Open the peel-pack. Take out the filter/opener with the tweezers.
Hold the CryoPen/CryoProbe's tip end pointed downward. Place the filter
in the instrument and make sure that it is seated flat at the bottom
of the loading cavity.
12
Empty loading cavity
Converter 16/8g: Adapter to
convert the backend
for a 8g cartridge
Blistered cartridge and
assembled filter/opener device

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis