Herunterladen Inhalt Inhalt Diese Seite drucken

Errors/Complications During The Calibration; Abort The Calibration - Dräger VarioGard 2300 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Maintenance
6.3

Errors/complications during the calibration

Action
Status display
Yellow/red flashes
alternately
Stop the calibration gas. Remove the calibration adaptor VarioGard 23x0.
Place the
magnetic wand for
2 seconds on the
marking »
«
for zero-point
Green flashes
calibration or
alternately
»
« for span
calibration and
then remove it
again.
Remove the
Green flashes
magnetic wand.
alternately
Possible corrective measures:
Repeat the calibration.
Check the calibration gas concentration.
The flushing time of the calibration gas is not sufficient.
If there has been several unsuccessful attempts, replace the
sensor.
16
Output
Meaning
signal
Error in
The device has
calibration
detected an error or
complications.
Maintenance
Error is
mode
acknowledged.
Maintenance
Measurement mode is
mode
activated with
previous setting for
zero-point or span.
6.4

Abort the calibration

Action
Status display
Stop the calibration gas. Remove the calibration adaptor VarioGard 23x0.
Place the
magnetic wand for
2 seconds on the
marking »
«
for zero-point
Yellow flashes
calibration or
»
« for span
calibration and
then remove it
again.
Green/yellow
flashes alternately
Remove the
Green/yellow
magnetic wand.
flashes alternately
Output
Meaning
signal
The instrument has
Maintenance
detected an abort
mode
command from the
user.
Maintenance
Instrument
mode
acknowledges the
abort.
Maintenance
Measurement mode is
mode
activated with
previous setting for
zero-point or span.
Dräger VarioGard 2300 / 2320

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Variogard 2320

Inhaltsverzeichnis