The screen will show S 232 or S USb. Press the [Mode] key to select which
interface you would like to use. Press [Tare] to confirm setting and configure
interface.
6) The following options are available for setting the output and accumulation
functions:
Mode
P1 Prt
P2 Con
P3 AUT
Press the [Mode] key to change the setting. Press [Tare] to confirm and move
to the next parameter.
7) Setting the baud rate. This is the transmission speed for communication with
printers and computers. The selected rate must match that of the other
device for communications to work.
The following settings are available:
b
b 1200
b 2400
b 4800
b 9600
Press the [Mode] key to change the setting. Press [Tare] to confirm and move
to the next parameter.
8) Parity settings. Parity is a communications check. There are 3 settings that the
Highland can work with as follows:
© Adam Equipment Company 2014
Print Feature
Data is sent whenever the [Print] key is
pressed.
Data is sent continuously.
The weighing results will be sent to the
communication port automatically
whenever a stable reading is present.
The balance has to return to zero
before another reading is sent via the
interface.
600
16
9520revF-Jun14
Accumulation
Manual accumulation when
the [Print] key is pressed.
Accumulation is disabled.
Automatic accumulation when
stable