Description Of User Interface; Keypad - Dentsply Maillefer wave one Gebrauchsanleitung

Endo-motor
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 58
6.4.

Description of User Interface

6.4.1. Keypad

2
1
1
POWER:
2
LED BATTERY:
3+4
:
5+6
:
7
OK:
8
CAL:
9
A.S.R.:
8
9
Switches the device on (LED on) and off (LED off).
Shows battery status; signals that the battery requires recharging
(for details see chapter 6.3.3).
Scrolls right/left in the bottom row of the display through the file
system, file, torque and speed fields. The active field is indicated
www.enversaltan.com
by an arrow (possible only for continuous rotary systems).
The + and – keys allow to scroll through the file systems and files
and to individually change the torque or speed values in all
continuous rotary file systems.
Confirms modifications to the torque and speed settings in each
system where modifications are possible. Also reloads the default
parameters when used together with the power key. The OK key
may also be used to start the micromotor without the foot pedal.
Calibrates the contra-angle to ensure torque accuracy each time
the contra-angle is replaced or lubricated.
Continuous rotation:
Green LED: Auto-stop-reverse motion followed by forward rotation
when preset torque is reached
Red LED: Reverse rotation without torque limitation, controlled by
foot pedal
LED Off: When the preset torque is reached, micromotor will stop
Reciprocating rotation:
A.S.R. is disabled (LED off).
www.enversaltan.com
3
5
7
6
Fig. 4 Keypad