Storage

When moving the PhysioStim device from very cold or very hot storage areas
(like your car), wait at least an hour to use or charge the device. The device
requires time to return to operating temperature.
Unpacked Storage:
Temperature Range:
• -25°C to 5°C
• 5°C to 35°C at up to 90% relative humidity, non-condensing
• 35°C to 60°C at a water vapor pressure up to 50 hPa
Packed Storage, Shipping and Transport:
Temperature Range: within -40°C to 60°C
• Between 10-100% relative humidity
• Including condensation at pressures between 500 hPa and 1060 hPa

Operating Environment:

Temperature Range: within 5°C to 40°C
• 15-90% relative humidity, non-condensing but not requiring a water
vapor pressure greater than 50 hPa
• 700-1060 hPa
The PhysioStim device is designed for a storage life of twelve months plus one
year of usage.

Travel

When traveling by air, it is recommended to pack the PhysioStim device with
checked luggage. If taken onboard the airplane, it should be turned off
when passing through security screening equipment, as the device could be
damaged. The PhysioStim device instruction manual should be taken with you
to quickly and easily identify the device for security personnel. Do not wear or
operate the PhysioStim device while onboard the airplane.

Disposal

After treatment is complete and a physician advises you to discontinue use,
you may dispose of the device according to your local governing ordinances or
recycling plans. Contact your local authorities to determine the proper method
for disposal since this is electronic equipment containing a lithium-ion battery.
You may also contact Orthofix Patient Services regarding recycling. The PhysioStim
device is for single patient use.
The PhysioStim device is a Class IIa medical device (prescription only) that
cannot be sanitized or used by another person.
Dispose of the device properly to prevent injury.
DO NOT dispose of the PhysioStim device in an incinerator. This
device contains lithium batteries.