Start-up
3
Insert the power supply cord into the docking station's 12 VDC port.
4
Remove the power supply cover.
5
Replace the cover with the adapter that is suitable for your outlet. If a dedicated power cord was ordered, use
it in place of the adapter.
6
Plug the power supply (or cord) into a suitable outlet.
7
Check the docking station display panel. When the unit has successfully completed its start-up process, the
green LED should be on and the text message should include:
"Cal Station" for Standalone Mode orders.
"√ iNet" for DSXi units and iNet Accounts. If the message doesn't display within 15 minutes, the unit is
not cloud-connected. DSXi users, recheck the work you did in "Activation". For iNet Accounts, contact
Industrial Scientific.
8
Fresh-air fitting
8
The fresh-air fitting is installed at the factory and is connected to the blue intake port labeled ZERO AIR. If
your application requires a zero-grade-air cylinder, remove the fitting: turn its white swivel connector
counterclockwise.
9–14
iGas cylinder connections
Supplies: compatible iGas cylinders equipped with demand-flow
regulators; iGas card-reader-and-tubing assemblies (assembly).
Connections for calibration gas are described and shown for
the port set iGAS 2−CAL GAS. If connecting a zero-grade-air
cylinder, follow the instructions below, but use the port set
iGAS 1−ZERO AIR
9
Attach the demand-flow regulator to the cylinder. Turn the cylinder clockwise to tighten.
10
Slide the iGas card, which is attached to the cylinder, into the assembly's card-reader.
11
Attach the open end of the assembly's tubing to the regulator nipple.
12
If the regulator is equipped with a pressure switch, plug its tab into the slot on the side of the card-reader.
13
Plug the assembly's cable connector (arrow faces up) into the iGAS 2 cable port.
14
Attach the tubing's white swivel connector to the cable port's corresponding, vertically aligned CAL GAS
intake port. Turn the connector clockwise to tighten.
Repeat steps 9-14 to connect additional calibration gas cylinders to the port sets, iGAS 3, 4, 5, or 6.
15-16
Data and settings
Standalone Mode orders
only
15
To automatically save bump test and calibration records (plus some instrument data) to the USB device,
connect it to the docking station's USB port. (Device not shown.) It is important to do this because no records
or data are ever saved to the docking station itself.
16
To ensure your USB-saved data are correctly date-and-time stamped, refer to the Product Manual (Setup
chapter). As instructed there, use the online DSX Configurator to set the station's date, time, and other
options. Note: After configuration, the Ethernet cable can be disconnected from the docking station.
.
Supplies: The USB data storage device that was included with your DSX
Standalone Mode unit.
Port set iGAS1
20
Port set
iGAS 2
CAL GAS
ZERO AIR