7.2 CLEANING THE OPTICS

Exercising proper care for the glass optic in your LED 4000 lightsource can help maintain
light quality and extend usable lifetime. Using the proper cleaning products and methods
are as important as cleaning the component itself.
Cleaning Solutions
There are a variety of cleaning solutions you can use such as Reagent-Grade Isopropyl
Alcohol, Reagent-Grade Acetone and De-Ionized Water to ensure a long product lifetime.
A good solvent mix to use is a mix of 60% acetone* and 40% methanol. (Acetone alone
dries too quickly to dissolve all of the debris. The methanol slows the evaporation time, and
also dissolves debris that acetone alone would not clean.)
*Acetone should never be used on plastic optics or components as it will damage them.
Optic Cleaning Procedure
Step #1
First use compressed air (aerosol can) to remove surface dust without directly contacting
any coating the optic may have. (Wiping a dusty optic with cotton swabs or lens paper can
abrade and damage the optic.)
Step #2
Soak with solvent a Cotton-Tipped Swabs or Lens Tissue then use a circular motion
starting from the edge of the optic moving towards the center to wipe away any dust or dirt.
Repeat as necessary, each time use a new cotton swab or lens tissue.
8. SERVICING AND REPAIR
Defective items are to be serviced and repaired exclusively by persons authorized by the
manufacturer. All repair work shall employ original manufacturer's parts only.
8.1 FUSE REPLACEMENT
Always disconnect power cord and turn main switch off before fuse
replacement.
CAUTION
Turn light source off and unplug power cord. On back of unit remove the fuse holder located
in the power cord receptacle and pull out fuse holder. Replace fuses with 3 AMP Slow-Blow
250V rated fuse. Insert holder back into fuse housing.
Re-connect the power cord and turn the LED light source on according to section 6.1.