Herunterladen Inhalt Inhalt Diese Seite drucken

Installation - Sunoptic Technologies LED-4000 Bedienungsanleitung

Led-lichtquelle
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31

5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable
gases.
DO NOT block the venting grids of the light source or impede exhaust air
flow. Users must ensure that ventilation air temperatures are below the
maximum rating for the light source and are not influenced by warm
exhaust from other sources.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light
source.
Use only cords provided with the light source.
CAUTION
Plug the supplied AC power cord into a wall outlet.
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings. Make sure that the power cord is connected
to the main power a suitable plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
5.2 CONNECTING THE FIBER OPTIC CABLE
Connect the light fiber optic cable to the instrument and then plug the light guide
endfitting into the appropriate turret port on the front panel.
NOTE
The turret has a light guide sensing feature that will not allow the LED light
to turn on until a fiberoptic cable is inserted into the active turret port.
LIT-216 SUNOPTIC TECHNOLOGIES
(English)
®
Date of Revision:01/27/16
Rev. -
Page 9 of 72

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis