6. OPERATION

6.1 POWERING UP THE LED LIGHT SOURCE
To operate the LED light source: Turn on the power switch. The indicator on power switch
will light.
6.2 LIGHT BRIGHTNESS DISPLAY AND CONTROL
The light brightness adjustment enables the user to obtain a brighter or darker illumination
of the object of observation.
Adjust the light intensity by turning the intensity control knob.

7. CLEANING AND DISINFECTION

NOTE
Cleaning and disinfection agents should be commercially available cleansers commonly
used for cleaning and/or disinfection of electronic equipment cabinets in hospitals, such as
ethyl or isopropyl alcohols, disinfecting sprays containing quaternary ammonium
compounds or hydrogen peroxide.
Do not use strongly caustic or acidic cleaners such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid or similar products. Do not use acetone, methyl ethyl ketones or
halogenated/chlorinated hydrocarbon solvents, or cleaners containing any of these
restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids over the
cabinet.
Do not allow liquids to enter the cabinet seams or ventilation openings, or remain within
switch or control device crevices.
8. SERVICING AND REPAIR
Defective items of equipment are to be serviced and repaired exclusively by persons
authorized by the manufacturer. All repair work shall employ original manufacturer's parts
only.

8.1 FUSE REPLACEMENT

CAUTION
Turn light source off and unplug power cord. On back of unit remove fuse by turning fuse
cover and pulling out fuse. Replace fuse with 2 AMP (1/4" x 1 ¼") time delay 250V rated
fuse. Insert back into fuse housing.
Re-connect the power cord and turn the LED light source on according to section 6.1.
LIT168 SUNOPTIC TECHNOLOGIES
(English)
Downloaded from
www.Manualslib.com
Always disconnect the power cord before cleaning and while unit is
drying if wet-wiped.
Always disconnect power cord and turn main switch off before fuse
replacement.
manuals search engine
®
Date of Revision:06/29/17
Rev. B
Page 10 of 60

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis