NOTE Both the high and low temperatures must be entered for a valid
calibration. If power is lost before the procedure is complete, critical data
needed for the calculation of calibration parameters will be lost. After
pressing
at the Stor prompt wait several seconds before proceeding
to ensure that a bad calibration message does not appear. Premature use
of the keypad after pressing
message.
CAL
rtd1
rtd1 = internal
sensor
r1H
r1H
r1L
r1L
may cancel the bad calibration error