Herunterladen Inhalt Inhalt Diese Seite drucken

R82 Toucan 3 Bedienungsanleitung Seite 18

Vorschau ausblenden Andere Handbücher für Toucan 3:
Inhaltsverzeichnis

Werbung

Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a
device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to main-
tain a device may compromise the clinical condition or safety of users and/or their caregivers.
EN
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any dam-
aged or worn-out parts
Ensure that all fasteners are in place and fit-
ted correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Weekly
Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
Wipe off any residue and dirt from the prod-
uct using a cloth wrung in warm water and
a mild detergent/soap without chlorine and
dry before use
Check that the angle mechanism functions
correctly and that the safety lock is working
correctly
Disinfect the product. Please check the
disinfection section in this User Guide for
more information
Monthly
All nuts and bolt on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substan-
dard or faulty
ENGLISH
Warewashing machine
This product can be washed at 60° using a
mild detergent for 15-20 minutes in a washing
machine designed for medical devices. Use the
machine's drying function to dry the product.
Remove all accessories before washing the main
product. Please allow the product to dry before
reuse.
Before washing remove the following items and clean
them separately:
Accessories
Cushioning
Disinfection
The product without accessories can be disinfected
with a 70% disinfectant IPA solution. It is recom-
mended to wipe off any residue and dirt from the
product, using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry before
disinfection.
Accessories and spare parts
The products from R82 can be supplied with a vari-
ety of accessories which comply with the needs of
the individual user. Spare parts can be ordered on
request. Find the specific accessories on our website
or contact your local dealer for further information.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in ac-
cordance with the manufacturer's instructions and
demonstrably recorded.
Presales information
Please go to support.r82.org
18
support.R82.org

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Toucan 4Toucan 2Toucan 1Toucan 5

Inhaltsverzeichnis