3. Technical details
Temperature display:
Menu mode indication:
Measurement range:
Power supply:
Power consumption:
Pyrometer interface:
(galvanically separated)
Operating Temperature:
Storage temperature:
Weight:
Enclosure rating:
Display resolution:
Housing:
Front panel cut-out:
Mounting depth required:
Connector type:
•
Measurement rate: dependent on Baud rate, e.g. approx. 50 ms at 19.2 kBd
•
Display update rate: 300 ms, or immediately (< 30 ms) when changes are > 5 K
•
Display switchable between °C / °F
•
Menu structure dependent on type of pyrometer connected and operating mode
4. Electrical connections
When a digital pyrometer is connected to the interface connection of the DA 6000-N, all parameters that are
usually set with a PC (e.g. with the InfraWin PC software), can be set using the parameter adjustments of
the digital display. The pyrometer is automatically identified when it is connected and the operating functions
of the connected device adapted. The power supply of a digital pyrometer cannot be the DA 6000-N.
[ Mes]
Mode:
7 segment LED, 13 mm, red
4-digit LED matrix, 5 mm, red
-100.0 to 3200 °C
(-148.0 to 5792 °F)
85 to 265 V AC, ripple max. 0,5 V
or:
18 to 30 V DC, 48 to 62 Hz
Approx. 7 VA (without external devices connected)
Either RS232 or RS485 half duplex
Baud rate 1.2 - 38.4 kBd
For communication with a digital pyrometer
0 to 50 °C on the housing
-20 to 70 °C
approx. 350 g
Front panel IP 40, otherwise IP 20 (DIN 40 050)
0.1 °C for temp. values up to 999.9 °C, 1 °C for temps. of 1000 °C and above
0.2 °F for temp. values up to 1831.8 °F, 1 °F for temps above that
Plastic housing (flame-retardant Noryl), 48 x 96 x 141 mm
(DIN 43 700) front panel fixed with screw clips
+0.6
+0.8
45
x 92
mm
≥ 165 mm with clips and cable
Plug-in clips, clamping area 0.13 - 2.5 mm² (AWG 28 - 16),
with conductor end tubes: 0.5 - 1.5 mm²
Digital
pyrometer
Interface cable
(RS232 or RS485)
pp
[ Mes]
DA 6000-N
11