18
This device uses the oscillometric method of blood pressure measurement.
This blood pressure monitor is intended for measuring diastolic and systolic
blood pressure and pulse rate among adults at home as well as in a medical
facility. To measure blood pressure, you have to attach the cuff to the upper arm
(recommended left) as described in the section "Attaching the cuff" and turn on
the device.
EN
• To prevent injuries, before taking measurement make sure that air tube is not
tangled.
• Do not measure more than 3 times at a time. Minimal interval between each
measurement is 5 minutes. Continuous measuring may cause bruising.
• Do not measure more than 6 times a day.
• Do not attach the cuff to the injured arm, as it may lead to further injuries.
• Do not measure blood pressure on the arm on the side of a mastectomy as it
may lead to further injuries.
• Observe the measurement progress from the display.
• If pressure in the cuff exceeds 280 mmHg; immediately turn off the device by
pressing the ON/OFF button.
• Do not bend the cuff or the air tube.
This device cannot be used by children below 12 and by people
who cannot express their intentions. If the device is used by people
between 12 - 18, make sure they are supervised by an adult.
Do not use the device for purposes different than measuring blood
pressure.
Do not use mobile phones when operating the device. Do not
expose the device to magnetic field.
Do not move during measurement.
Do not disassemble and repair the main unit and the cuff. Risk of
electric shock and damaging the device.
INTENDED USE
NOTES
Owner's manual