mAINTENANcE, cARE, AND SERvIcE

Order all spare parts from your dealer or pharmacist, or contact Colpharma-Service (see foreword).
We recommend replacing the nebuliser, mouthpiece, and the face masks after 3 months use.
Check the filter continually for cleanliness and replace it if dirty, or after a maximum of 3 months use.
Spare filters are provided with the device.
To replace the filter, open the air filter compartment (4) on the top of the device marked «Filter» and exchan-
ge the filter (10).
Never unscrew the plastic lid.

mALFUNcTIONS AND AcTIONS TO TAKE

The device cannot be switched on
Ensure the power lead (2) is correctly plugged into the socket.
Ensure the ON/OFF switch (3) is in the position «I».
The nebuliser functions poorly or not at all
Ensure the air hose (6) is correctly connected at both ends.
Ensure the air hose is not squashed, bent, dirty or blocked. If necessary, replace with a new one.
Ensure the nebuliser (5) is fully assembled and the colored vaporiser head (11) is placed correctly.
Ensure the required medication has been added.